4 things to know about pedicle screw designs in spine surgery

Dual-threaded pedicle screws provided the strongest primary pull-out strength for spinal fixation, according to a study published in Spine.

Advertisement

The researchers tested the pull-out strengths of three different pedicle screw designs (single-thread, mixed-thread, dual-thread) in standardized polyurethane foam in three sequences.

Four things to know:

1. The primary pull-out strength was similar across screw designs, although dual-thread screws showed higher primary pull-out strength (2628.8 newtons) compared to single-thread screws (2184.4 newtons).

2. For revision pull-out strength, the mixed-thread screws had an 18.6 percent reduction in pull-out strength. Revision with a larger diameter screw improved the pull-out strength back to baseline.

3. The researchers found no significant reduction in revision pull-out strength for single and dual-thread screws.

4. In cases in which mixed-threaded screws must be revised (at the index or revision surgery), surgeons should consider using larger diameter screws to improve the pull-out strength, the study authors concluded.

More articles on devices:
Medtronic Q3 spine revenue flat at $655M, Mazor X & StealthStation help drive neurosurgery growth: 6 things to know
Customized knee implants cost less than standard implants — 5 study insights
Integra LifeSciences launches nail system for ankle fusion: 3 things to know

At the Becker’s 32nd Annual Meeting: The Business and Operations of ASCs, taking place October 29-31 in Chicago, ASC leaders, surgeons and healthcare executives will explore strategies to drive growth, enhance operational performance, navigate reimbursement challenges and prepare for the future of ambulatory surgery. Apply for complimentary registration now.

Register to Attend Webinar

160 ambulatory leaders just ranked the EHR as the single system most overdue for AI reinvention

Tuesday, August 11
12:00 PM - 1:00 PM CDT

Presenter: Gautam Shah, MBA, FACHDM, NextGen Healthcare

Advertisement

Next Up in Spinal Tech

Advertisement

Comments are closed.